This probe is capable of measuring pulse currents up to 100 amperes with wide operating frequency of 10 KHz - 500 MHz.
Conducted currents can be measured without making direct contact with the source conductor or metallic surface by means of clamp-on current probes. The BCP-615 Current Probe is designed to permit field intensity meters, spectrum analyzers, and other 50 ohm impedance instruments to measure quantitative magnitudes of current. Measurements can be made on single and multi-conductor cables, ground and bonding straps, shielded conduits and on coaxial cables. A current probe acts as a single turn primary, multiple turn secondary transformer, placing low series impedance in the probed power line or signal lead while capable of driving a usable signal into a 50ohm receiver. A current probe is characterized by its transfer impedance, ZT (dBohm) the ratio of output voltage into a standard load (usually 50ohm for EMI testing) divided by the net current traveling through the probe window area.
For ease and convenience of performing conducted measurements, all of our current probes utilize the split type clamp-on design. Small and lightweight, each Current Probe is manufactured to exacting standards, thus insuring repeatable performance.
Recommended for RTCA DO-160E/F section 20
Frequency Range: | 10 KHz - 500 MHz |
Transfer Impedance (dBohm) | -40 to 18 |
Max Primary Current (DC - 60 Hz) |
400 Amps |
Max Primary Current (400 Hz) |
350 Amps |
Max Primary Current CW | 80 Amps |
Peak Pulse Current | 100 Amps |
Connector: | N-Type, Female |
Physical Dimensions | |
Aperture: | 1.25" (32 mm) |
Weight: | 1.2 lb.'s (0.55 Kg) |
A Basic Bulk Current Injection Test Method