Product Description
The ICP-621 injection current probe is used for specifications that require the injection of large frequency currents into cable bundles and individual wires. It can be used as an injection source as well as a sensitive monitoring probe. The ICP-621 has an input power rating of 100 watts CW for 30 minutes and covers the frequency range of 10 KHz - 100 MHz.
An injection current probe acts as a multiple turn primary and a single turn secondary transformer, when placed around a power line or signal lead. Thus it provides a (nominal) 50ohm load to the susceptibility signal source, while providing a lower susceptibility signal source impedance when placed in series with the cable under test. An injection current probe is characterized by its insertion loss (dB). The insertion loss describes the inefficiency of the clamp relative to direct injection into a 50ohm circuit.
Specifications
| Frequency Range: |
10 KHz - 100 MHz |
| Insertion Loss |
8 to 30 dB |
| Max Input Power |
100 watts |
| Max Input Current |
26 Amps |
| Max Time at Full Power |
30 minutes |
| Max Temperature Rise |
35° C |
| Connector: |
N-Type, Female |
| Physical Dimensions |
|
| Aperture: |
1.5" (38 mm) |
| Diameter: |
4.25" (108 mm) |
| Weight: |
4.2 lb.'s (1.9 kg) |
The information provided below is for general informational purposes only and should not be relied upon for any specific purpose.
Current Probe Fixture
CPF-630
Injection Probe calibration fixtures are a part of the test equipment required by most of the Bulk current Injection Test Procedure Specifications. The Calibration Fixture allows the user to quickly and easily calibrate the injection probe prior to performing the compliance testing. It exceeds DO-160, Section 20 VSWR fixture requirements for stand-alone injection probe calibration fixture.
Product Details
3 Meter N/N Cable
SAC-213
3 Meter RF cable with N(m) to N(m) connectors. Custom lengths can be made to suit your specifications.
Product Details